For full functionality of this site it is necessary to enable JavaScript.

HIOKI 3561 电池测试仪

Specification
- Max. applied measurement voltage:
  + ±22 V DC rated input voltage
  + ±70 V DC maximum rated voltage above ground
- Resistance Measurement ranges
  + 300mΩ(max.310mΩ/resolution 10μΩ)~3Ω (Max.3.1Ω/ resolution  100μΩ), 2 range/resolution : ±0.5 % rdg. ±5 dgt
  + Resolution :±0.5 % rdg. ±10 dgt
  + Testing sound Frequency: 1kHz ±0.2Hz
  + Testing current: 10mA (Thang 300mΩ)~1mA (Thang 3Ω)
  + Open-circuit Votalge: 7V
- Voltage Measurement ranges:     DC 20 V(resolution 1mV)DC
- Display: 31000 full digits (resistance), 199999 full digits (voltage), LED
- Sampling rate: Four steps:4 ms(Extra-FAST), 12 ms(FAST), 35 ms(Medium), 150 ms(Slow)
- Measurement time: 3ms
- Comparator functions:
  + Setting: Upper and lower limit
  + Output: 3 levels (Hi, In, Lo) 
- Analog Output:      Measured resistance (displayed value, from 0 to 3.1V DC
- Interface: External I/O, RS-232C, Printer (RS-232C)
- Power supply: 100 to 240 V AC, 50/60 Hz, 30 VA max
- Dimensions, mass: 215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.4 kg (84.7 oz)
- Supplied accessories: Power cord × 1
- The warrnty period: 12 month

Details
Max. applied measurement voltage ±22 V DC
±70 V DC maximum rated voltage above ground
Resistance Measurement ranges 300 mΩ (Max. display 310.00 mΩ, resolution 10 μΩ) to 3 Ω (Max. display 3.1000 Ω, resolution 100 μΩ), 2 ranges
Accuracy: ±0.5 % rdg. ±5 dgt. (Add ±3 dgt. for EX.FAST, or ±2 dgt. for FAST and MEDIUM)
Testing source frequency: 1 kHz ±0.2 Hz, testing current: 10 mA (300 mΩ range), 1 mA (3 Ω range)
Open-circut Voltage: 7 V peak7
Voltage Measurement ranges DC 20 V, resolution 0.1 mV
Accuracy: ±0.01 % rdg. ±3 dgt. (Add ±3 dgt. for EX.FAST, or ±2 dgt. for FAST and MEDIUM)
Display 31000 full digits (resistance), 199999 full digits (voltage), LED
Sampling rate
Four steps:   4 ms(Extra-FAST), 12 ms(FAST),
35 ms(Medium), 150 ms(Slow)
(Typ., sampling time depends on supply frequency settings and function.)
Measurement time Response time + sampling rate, approx. 3 ms for measurements
(Response time depends on reference values and the measurement object.)
Comparator functions Dual comparators independently judge Hi/IN/Lo states of resistance and voltage
Setting: Upper and lower limit, Deviation (%) from a standard value
Output: 3 levels (Hi, In, Lo), besides independent judgment results for resistance and voltage, ANDed calculation results
Display, Dual audible indicato, or external I/O terminal, Open-collector (35 V, 50 mA DC max.)
Interface External I/O, RS-232C, Printer (The 9670 via RS-232C), GP-IB (-01 suffix models only)
Power supply 100 to 240 V AC, 50/60 Hz, 30 VA max.
Dimensions, mass 215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.4 kg (84.7 oz)
Supplied accessories Power cord ×1