Hioki 3169-20电能质量分析仪

SPECIFICATIONS
- Measurement line type: 
   + Single-phase 2-wire, single-phase 3-wire
   + Three-phase 3-wire, three-phase 4-wire
   + Voltage 4 channels, Current 4 channels
- Measurement items: 
   + Voltage, Current, Active/reactive/apparent power
   + Active/ reactive power integration 
   + Power factor, Frequency, Harmonic waveform value
- Measurement range:
   + 150~600V AC, 3 ranges
   + 500mA to 5 kA AC
   + Power range: 75W to 900kW 
- Basic accuracy:
   + AC Voltage: ±0.2 % rdg. ±0.1 % f.s
   + AC Current: ±0.2 % rdg. ±0.1 % f.s
   + Active power: ±0.2 % rdg. ±0.1 % f.s 
- Data save interval: 1~30s, 1~60min, 13 selects
- Other functions:
   + Error connect check, language selection, display hold
   + Setting backup, power shut off management, key lock
- External interfaces:
   + RS-232C
- Power supply: 100 to 240 V AC, 50/60 Hz, 30 VA max
- Product Net Weight: 1.2 kg
- Product Size: 210*160*60mm
- Supplied accessories:
   + VOLTAGE CORD L9438-53 ×1 (Black/ Red/ Yellow/ Blue)
   + Power cord ×1, Instruction Manual ×1, Quick Start Manual ×1
   + CD-R ×1 (RS-232C interface operating manual) ×1
   + Input cord label ×1
- The warranty period: 12 month
Details

 

Measurement line & number of circuits Single-phase 2-wires (4 circuits), Single-phase 3-wires (2 circuits), Three-phases 3-wires (2 or 1 circuit), and Three-phases 4-wires (1 circuit)
Note: 50 or 60 Hz, and that share the same voltage line
Measurement items Voltage, Current, Active/ reactive/ apparent power, Active/ reactive power integration, Power factor, Frequency,
Harmonic waveform value (up to 40th order)
Voltage ranges 150.00 V to 600.00 V AC, 3 ranges
Current ranges 500.00 mA to 5.0000 kA AC (depends on current sensor in use)
Power ranges 75.000 W to 900.00 kW (depends on combination of voltage, current range, and measurement line)
Basic accuracy AC Voltage: ±0.2 % rdg. ±0.1 % f.s.
AC Current: ±0.2 % rdg. ±0.1 % f.s. + current sensor accuracy
Active power: ±0.2 % rdg. ±0.1 % f.s. + current sensor accuracy (at power factor = 1)
CLAMP ON SENSOR 9661: ±0.3 % rdg. ±0.01 % f.s. (different from each model)
Measurement method Digital sampling, PLL synchronization or 50/60 Hz fixed clock
Basic accuracy Active power: ±0.2% rdg. ±0.1% f.s. + clamp-on sensor accuracy (at 45 to 66 Hz)
Display refresh rate 2 times /sec (except when using a PC card while accessing the internal memory, or when performing RS-232C communications)
Data save interval Standard interval: 1 sec to 30 sec, 1 minute to 60 minutes, 13 selects
Fast interval: A single waveform, or 0.1, 0.2, or 0.5 sec (at instant value only)
Frequency characterristics Fundamental waveforms up to the 50th order ±3% f.s. + measurement accuracy (of a 45 to 66 Hz fundamental waveform)
Other functions Error connect check, language selection, display hold, setting backup, power shut off management, key lock, [3169-21 only] D/A output 4 channels, ±5 V DC f.s.)
Power supply 100 to 240 V AC, 50/60 Hz, 30 VA max.
Dimensions, mass 210 mm (8.27 in) W × 160 mm (6.30 in) H × 60 mm (2.36 in) D, 1.2 kg (42.3 oz)
Supplied accessories VOLTAGE CORD L9438-53 ×1 (Black/ Red/ Yellow/ Blue), Power cord ×1, Instruction Manual ×1, Quick Start Manual ×1, CD-R ×1 (RS-232C interface operating manual) ×1, Input cord label ×1, CONNECTION CABLE 9441 ×1 (for the 3169-21 only)
 

OPTION

Voltage measurement
VOLTAGE CORD L9438-53
Bundled with standard, Black/ Red/ Yellow/ Blue, 3m (9.84ft) length, Alligator clip ×4
MAGNETIC ADAPTER 9804-01
Attaches to the tip of the Voltage Cord, Red ×1
MAGNETIC ADAPTER 9804-02
Attaches to the tip of the Voltage Cord, Black ×1
Current measurement
For power or load current measurement (1 sensor necessary for single-phase measurements, and 2 or 3 sensors required for 3-phase measurements)
CLAMP ON SENSOR 9694
CLAMP ON SENSOR 9661 5A AC rated current, φ 15 mm (0.59 in), 3 m (9.84 ft) length
 
Details
CLAMP ON SENSOR 9660
CLAMP ON SENSOR 9660 100A AC rated current, φ 15 mm (0.59 in), 3 m (9.84 ft) length
 
Details
CLAMP ON SENSOR 9661
CLAMP ON SENSOR 9661 500A AC rated current, φ 46 mm (1.81 in), 3 m (9.84 ft) length
 
Details
FLEXIBLE CLAMP ON SENSOR CT9667
CLAMP ON SENSOR 9661 5000/500 A AC rated current, φ 254 mm (10 in), Cable length: Between sensor - box 2 m (6.56 ft), Output cable 1 m (3.28 ft)
 
Details
CLAMP ON SENSOR 9669
CLAMP ON SENSOR 9661 1000A AC rated current, φ 55 mm (2.17 in), 3 m (9.84 ft) length
 
Details
CLAMP ON SENSOR 9695-02
50A AC rated current, φ15mm(0.59in), Requires the Connection Cord 9219
CLAMP ON SENSOR 9695-03
100A AC rated current, φ15mm(0.59in), Requires the Connection Cord 9219
CONNECTION CORD 9219
Connect with the 9695-02/ -03, Output BNC terminal
For clamp sensor adapter

 

CLAMP ON ADAPTER 9290-10
CT for 1000A AC, secondary current 1/10 of primary
 
Details
DOWNLOAD CATALOG
pdficon 3169-20, 3169-21
product catalog.pdf (1.41 MB) (2012-09-06 updated)

 

 