For full functionality of this site it is necessary to enable JavaScript.

Fuootech FTCT-1000F 600A CT PT Tester Current Transformer Test Equipment (0 ~ 2500Vrms, 5kVA)

Operational Power Supply: AC220V±10%, 50Hz

Single Machine Output: 0~2500Vrms, 5kVA (maximum setting value 20A)

Primary Current Output: 0~1000A

Excitation accuracy: ≤0.5% (0.2%*reading+0.3%*range)

Secondary winding Resistance measurement:

Range: 0.1~300Ω

Accuracy: ≤0.5% (0.2%*reading+0.3%*range)

Secondary real Load measurement:

Range: 5~500VA

Accuracy: ≤0.5% (0.2%*reading+0.3%*range) ±0.1VA

Phase measurement (Angle difference):

Accuracy: 4min

Resolution: 0.1min

CT ratio measurement range:

Range: ≤25000A/5A(5000A/1A)

Accuracy: ≤0.5%

PT ratio measurement range:

Range: ≤500KV

Accuracy: ≤0.5%

Operating Temperature: -10ºC~+40ºC

Relative Humidity: <90RH%

Altitude: ≤1000m

Overall Dimension (Length×Width×Height): 410mm*260mm*340mm

Weight: ≤22 kg